How should I extract or purify my RNA samples if I wish to reverse transcribe them using EpiScript RT PLUS?

AS3107, MBTOOL-007, PCRAMP-104

MasterPure™, sbeadex™ and QuickExtract™ isolation chemistries are all suitable for use with EpiScript RT PLUS.

Which RT enzyme (MMLV, EpiScript RNase H- Reverse Transcriptase, or EpiScript RT PLUS) should I use for pathogen detection, gene expression, or RNA-Seq library preparation, and why?

AS3107, MBTOOL-007, PCRAMP-104

All three reverse transcriptases; standard MMLV, EpiScript RNase H- Reverse Transcriptase and EpiScript RT PLUS can be used for these applications. EpiScript RNase H- Reverse Transcriptase has the advantage of being active up to 55 °C making it more specific than standard MMLV RT enzymes. EpiScript RT PLUS provides the added benefit of allowing room temperature reaction setup.

What is the difference between EpiScript™ RNase H- Reverse Transcriptase and EpiScriptT RT PLUS?

AS3107, MBTOOL-007, PCRAMP-104

EpiScript RT PLUS contains a blocking mechanism that suppresses reverse transcriptase activity at lower temperatures allowing for room temperature reaction setup.

Why is my reverse transcription reaction not creating cDNA?

307

  • Primer issues
    • The gene-specific primer design is incompatible with high reaction temperatures
    • The gene-specific primer is not complementary to the target of interest.
    • Non-specific primers were used with a Tm that is too low
  • The incorrect reaction temperature was used.
  • Contaminants in the sample preparation interfered with the reaction conditions.
  • The reaction buffer conditions are not correct. Use the kit-provided 10X Thermostable RT Buffer provided with the kit.
  • How should I extract or purify my RNA samples if I wish to reverse transcribe them using EpiScript RNase H- Reverse Transcriptase?

    221, RDXERT300

    MasterPure™, sbeadex™ and QuickExtract™ isolation chemistries are all suitable for use with EpiScript RNase H- Reverse Transcriptase.

    What is the optimum temperature for EpiScript™ RNase H- Reverse Transcriptase?

    221, RDXERT300

    The optimum temperature for EpiScript RNase H- Reverse Transcriptase is dependent on a range of factors. These include the target sequence and structure, the type of reverse transcriptase (RT) primer that you are using, and whether you are doing 1-step or 2-step RT-qPCR.

    If your target sequence has a high %GC content or a high secondary structure, higher temperatures of 50 °C or 55 °C are better. If random primers are being used in the RT reaction, we do not recommend using temperatures above 42 °C as many hexamers and nonamers will not anneal at 50 °C or 55 °C.

    In standard 1-step RT-qPCR experiments (non-challenging targets), we found minimal differences in performance when using either 42 °C, 50 °C and 55 °C for incubation. In 2-step RT-qPCR, however, we noticed that higher temperatures for the RT reaction resulted in reduced Cq values.

    What is the recommended concentration for EpiScript RNase H- Reverse Transcriptase in RT reactions?

    221, RDXERT300

    For 1-step RT-qPCR, we have found 20-50 units per reaction to work best. In 2-step RT-qPCR, we have found 5-20 units per reaction to work best.

    What conditions do you recommend for set up of RT reactions using EpiScript RNase H- Reverse Transcriptase?

    221, RDXERT300

    For best results we recommend preparing reactions on ice for both 1-step and 2-step RT-qPCR experiments.

    Can I use random hexamer or nonamers with EpiScript RNase H- Reverse Transcriptase?

    221, RDXERT300

    Random primers can be used with EpiScript RNase H- Reverse Transcriptase. If using random primers, we suggest that the RT step is performed at 42 °C or lower as many random primers will not anneal at higher temperatures.

    What is the optimum temperature for Lyo-Ready EpiScript™ Thermostable Reverse Transcriptase?

    30260-1

    The optimum temperature for Lyo-Ready EpiScript™ Thermostable Reverse Transcriptase is dependent on a range of factors. These include the target sequence and structure, the type of reverse transcriptase (RT) primer that you are using, and whether you are doing 1-step or 2-step RT-qPCR.