Technical specs
Cq Uniformity
Users can expect minimal well-to-well variability in Array Tape, providing The IntelliQube demonstrated a total standard deviation (SD) of 0.087 cycles when 768 replicate wells each containing 2048 copies of a NIST calibrated human – gDNA reference standard were analyzed for a non- transcribed locus in hgDNA.

Figure 1. Cq Uniformity <0.09 SD
Dynamic Range
The IntelliQube can support assays across a large range of input concentrations while maintaining data integrity. When analyzing a 1000 bp synthetic target serially diluted across eight orders of magnitude, the IntelliQube demonstrated a linear dynamic range of quantifications between 4x107 to 40 copies/reaction.

Figure 2. Linear dynamic range of quantification from 4x107 to 40 copies/reaction
Resolution
Minimal differences in target quality can be accurately resolved between reaction wells on the IntelliQube. Using NIST calibrated hgDNA reference standard, reactions were prepared to represent 1.2 fold differences in template concentration. When analyzed for a non-transcribed locus in hgDNA, the IntelliQube demonstrated the ability to resolve 1.2 fold differences with greater than 97% sensitivity and specificity.

Figure 3. 1.2 fold resolution
Sensitivity
The IntelliQube provides the sensitivity necessary to amplify and detect even 1-2 copies of target DNA in a reaction well. Although the theoretical limit of detection (LOD) provided by Poisson statistics is limited to three molecules, IntelliQube has demonstrated the ability to detect as little as 1-2 molecules of hgDNA per reaction.
IntelliQube Performance Specifications
Dimensions (W x D x H) | 49" x 34" x 75" (with cart) |
---|---|
Weight | 855 pounds |
Input Voltage | 120/240 VAC, 15/7.5 A |
Frequency | 50-60 Hz |
Operating Temperature | 15-30°C |
Operating Humidity | 20-80% Relative Humidity, non-condensing |
Air Input | 80-100 psi, bursts up to 15 SCFM |
Water Input | Reverse Osmosis or Deionized |
Certifications | Conforms to the European Machinery Directive (2006/42/EC) and the EMC Directive (2004/108/EC). For Research Use Only. Not for use in Diagnostic Procedures. |
Estimated End-point PCR Throughput Per 8 Hours (With Hydrocycler2) | Sixty Five 384-well arrays 24,960 data points |
Temperature uniformity
Specification | Peltier block type |
---|---|
Block configuration | 384/768-well Array Tape |
Temperature range | 20 - 100°C |
Temperature accuracy | ±0.25°C |
Temperature uniformity | ±0.5°C at 60°C |
Heating ramp rate | 3.0°C/sec |
Cooling rate | 3.0°C/sec |
Detection
LEDs (460-645 nm) |
![]() |
||||
---|---|---|---|---|---|
Dye channel | 1 | 2 | 3 | 4 | 5 |
Excitation filter (nm) | 480 | 530 | 548 | 580 | 620 |
Emission Filter (nm) | 510 | 565 | 580 | 625 | 705 |
Dyes with expected compatability | FAM™ |
JOE™
VIC®/HEX™
Cal Fluor®
Orange 560 |
NED™
TAMRA™
Cal Fluor® Red 590
|
ROX™
Cal Fluor® Red 610 Texas Red®
|
Cy5®
Quasar® 670
|